Adjustable Thread Calibrations

ATS has the capability to calibrate AGD (American Gage Design) and Southern Style Thread Ring Gages. ATS has hundreds of Thread Setting Plugs with standards and special Pitch Diameter in English and Metric. We are also capable of calibrating special gages when a Setting Plug is unavailable from 0.12 inch to 5.5 inch PD.

ATS takes special care of cleaning and inspecting our customer’s gage before calibration to ensure the correct fit.

The Standard Lab is environmentally-controlled for high precision measurement capabilities. The lab is controlled to 68 degree F within a span of +/-1.0 degree F and humidity controlled at 30-55% R.H. The lab also has isolation joints in the flooring to reduce any vibration effects. All gages are cleaned prior to testing and are placed in the Standards Lab to stabilize to temperature for a minimum of 24 hours. Applied Technical Services participates regularly in “Round Robin Gage Measurement Proficiency Testing” in accordance with our A2LA Accreditation.

All Calibrations performed include AS Found and A Left Data and are traceable to NIST, The National Institute of Standards and Technology and are performed in accordance to documented and controlled procedures. Our Quality Assurance system complied with the applicable requirements of ANSI/NCSL, Z540-1, ISO 9001-2000, 10CFR50 Appendix B, 10CFR Part 21 and ISO.IEC 17025. Upon request, we can provide calibrations that are accredited to A2LA. You can also be set up on our web based interface, I portal, which gives you access to electronic copies of Calibration Certifications, current asset status, asset history and due date reports for your equipment.